SentroPAT BU II
High performance NIR analyzer for pharmaceutical powder blending | |
|
Main features
- High performance spectrometer engine, easy to mount
- accurate and reliable NIR technology
- Seamless method transfer between instruments
- GmP friendly with smooth surfaces
- Superior spectroscopic performance
- Low cost of ownership
- Hot swappable battery
- No data loss at communication failure
- OPC connectivity
- SiPAT interface
|
| Bin Blending Sentronic PAT solutions are widely used in pharmaceutical powder blending to increase process understanding and support process and product release |
Key benefits
- Enables process understanding for "black box" process of powder blending.
- Seamless method transfer between instruments
- Full integration support (mechanical, IT, processes)
- Low cost of ownership (maintenance free)
|
Technical Data | SentroPAT Blend Uniformity TL |
---|
Wavelength range | 1350 - 1800nm |
Spectral resolution | >= 0.3nm (configurable), specification at 2nm |
Wavelength accuracy | < 1nm |
Wavelength reproducibility | < 0.1nm |
Photometric linearity | Fulfilling USP <1119> requirements |
Noise - High flux | < 0.3 x 10-³ AU according to USP<1119> |
Noise - Low flux | < 1 x 10-³ AU according to USP<1119> |
Data acquisition time | 400msec |
Sampling interface | Non contact, through sapphire window |
Measurement spot size | 25mm uniform beam shape for homogeneous sample illumination effective detection area 12mm diameter |
Working distance | Depending of sapphire window (27mm at 8.5mm sapphire thickness) |
Power supply | 2 hot swappable batteries with in total > 5hours operation External power supply for lab operation and recharging |
Enclosure | IP65 / NEMA4, cGMP conform |
Dimensions | 220 x 210 x 150mm (W x D x H) Width: add 80mm for handle Height: add 50mm for flange |
Weight | 6.5Kg / 14.5lb |
Measurement trigger | Internal software controlled 3D position measurement |
Communications | Wireless LAN (802.11 b/g/n) |
Software | SentroSuite GmP |
Validation | Software integrated validation routines according USP<1119>, EP2.2.40 |
Service port | Connections for DVI-D monitor and 3x USB |
Mounting | 4" hygienic flange |